Experimental Study of a New Method for Early Detection of Vascular Access Stenoses: Pulse Pressure Analysis at Hemodialysis Needle KU Leuven Ghent University
Hemodialysis vascular access (VA) stenosis remains a frequent complication. However, early detection is challenging and costly. The aim of this in vitro study was to assess a new detection method based on pulse pressure analysis at the hemodialysis needle. A silicon model of a radiocephalic arteriovenous fistula was built in a mock loop. Pressure profiles were measured at the arterial hemodialysis needle and in the proximal feeding artery. ...