< Terug naar vorige pagina

Publicatie

Reliability, validity and clinical usability of a robotic assessment of finger proprioception in persons with multiple sclerosis

Tijdschriftbijdrage - Tijdschriftartikel

Background: Multiple sclerosis often leads to proprioceptive impairments of the hand. However, it is challenging to objectively assess such deficits using clinical methods, thereby also impeding accurate tracking of disease progression and hence the application of personalized rehabilitation approaches. Objective: We aimed to evaluate test-retest reliability, validity, and clinical usability of a novel robotic assessment of hand proprioceptive impairments in persons with multiple sclerosis (pwMS). Methods: The assessment was implemented in an existing one-degree of freedom end-effector robot (ETH MIKE) acting on the index finger metacarpophalangeal joint. It was performed by 45 pwMS and 59 neurologically intact controls. Additionally, clinical assessments of somatosensation, somatosensory evoked potentials and usability scores were collected in a subset of pwMS. Results: The test-retest reliability of robotic task metrics in pwMS was good (ICC=0.69-0.87). The task could identify individuals with impaired proprioception, as indicated by the significant difference between pwMS and controls, as well as a high impairment classification agreement with a clinical measure of proprioception (85.00-86.67%). Proprioceptive impairments were not correlated with other modalities of somatosensation. The usability of the assessment system was satisfactory (System Usability Scale ≥73.10). Conclusion: The proposed assessment is a promising alternative to commonly used clinical methods and will likely contribute to a better understanding of proprioceptive impairments in pwMS.
Tijdschrift: Multiple Sclerosis and Related Disorders
ISSN: 2211-0348
Volume: 70
Jaar van publicatie:2023
Trefwoorden:Multiple sclerosis, Assessment, Rehabilitation, Somatosensory function, Proprioception
Toegankelijkheid:Open