< Terug naar vorige pagina

Publicatie

Comparison of Immunoassays for Measuring Serum Levels of Golimumab and Antibodies Against Golimumab in Ulcerative Colitis: A Retrospective Observational Study

Tijdschriftbijdrage - Tijdschriftartikel

BACKGROUND: Golimumab is a monoclonal anti-tumor necrosis factor alpha antibody, which is used in ulcerative colitis with an exposure-response relationship. The goal of this study was to compare results obtained with different immunoassays (golimumab and antigolimumab antibodies trough levels). METHODS: This study was based on samples from 78 ulcerative colitis patients on golimumab treatment. Golimumab was quantified by either an anti-IgG detection antibody (Theradiag, Marne la Vallée, France) or an antibody directed against golimumab (Sanquin, Amsterdam, The Netherlands, KU Leuven, Leuven, Belgium, and Janssen R&D, San Diego, CA). Bridging drug-sensitive enzyme-linked immunosorbent assays (Theradiag, Janssen R&D, and KU Leuven), a bridging drug-tolerant enzyme-linked immunosorbent assay (Janssen R&D), and a radioimmunoassay (Sanquin) were used to quantify antidrug antibody. RESULTS: Median serum golimumab levels were 4.5, 3.5, 4.9, and 2.4 mcg/mL with Theradiag, Sanquin, KU Leuven, and Janssen R&D assay, respectively (P < 0.05). Correlation coefficients between assays ranged from 0.9 to 0.97. When using the KU Leuven and Janssen R&D assays, 86% of samples were in the same quartile of distribution of values, and for Sanquin and Janssen R&D assays, this overlap was 80%. The concordance observed for the other pairs was 83% (Sanquin/KU Leuven R&D), 71% (Theradiag/KU Leuven), and 68% (Theradiag/Janssen R&D and Theradiag/Sanquin). The specificity of assays for golimumab was demonstrated. Antidrug antibodies were detected in 28.2% of the samples with the Janssen R&D drug-tolerant assay and in the same 2 patients by the 3 other assays. CONCLUSIONS: Performances of these immunoassays were similar in terms of quality, but differences in the quantitative results point to the importance of using the same assay consistently to monitor a patient's treatment.
Tijdschrift: Therapeutic Drug Monitoring
ISSN: 0163-4356
Issue: 4
Volume: 41
Pagina's: 459 - 466
Jaar van publicatie:2019
BOF-keylabel:ja
IOF-keylabel:ja
BOF-publication weight:1
CSS-citation score:1
Auteurs:International
Authors from:Higher Education