< Terug naar vorige pagina

Publicatie

Fast and Accurate Identification of Dermatophytes by Matrix-Assisted Laser Desorption Ionization–Time of Flight Mass Spectrometry: Validation in the Clinical Laboratory

Tijdschriftbijdrage - Tijdschriftartikel

The performance of a matrix-assisted laser desorption ionization–time of flight mass spectrometry (MALDI-TOF MS) workflow using an extensive reference database for dermatophyte identification was evaluated on 176 clinical strains. Using a direct-deposit procedure after 3 incubation days yielded 40% correct identification. Both increasing incubation time and using an extraction procedure resulted in 100% correct identification.
Tijdschrift: J Clin Microbiol
ISSN: 0095-1137
Issue: 9
Volume: 52
Pagina's: 3440-3443
Jaar van publicatie:2014
Trefwoorden:Maldi-tof
  • WoS Id: 000341359300050
  • Scopus Id: 84906886296